#be sure to be in double res mode; 200MHz lowpass filter #need to spin at 20kHz etc as program is rotor triggered rpar hahnecho_dec.1p9 all pulprog dan_echo_dec ############################## # 13C SETTINGS # ############################## p1 2.5u #C90 p2 5.0u #C180 plw1 65 #C90/plw (for 100kHz RF) cnst31 20000 #spin rate 1 swh 60k #swh 3*vr or 4*vr etc 1 aq 10m #aq time o1p 100 #C center ppm ############################## # 1H SETTINGS # ############################## o2p 4.5ppm #H center plw13 13.4 #CW decoupling power during echo pulses plw12 13.4 #tppm15 decoupling power during acq pcpd2 5.6u #tppm15 pulse length cpdprg2 tppm15 #tppm15 dec program ############################## digmod baseopt #sets digitizer for solids nmr ns 8k rg 256 ds 0 d1 8s ############################## 1 SI 2k # processing points 1 PHC0 20.0 # direct phase0 correct 1 PHC1 0.0 # direct phase1 correct 1 WDW GM 1 GB 0.04 1 LB -40 ############################## ############################## ##############################