750 Helium Fill
Order 250 liter Helium for ~ 1 week delivery after we hit 60% on meter.
Fill magnet completely with N2(li) before doing a fill. Ice jams form inside ports if not done.
We need to slowly let the Helium overhead pressure drop to ~980 mBar before opening magnet. on top
of console is a meter which reads this pressure ("atmospheric pressure device"). Overhead pressure is usually 1.029 bar (1.0155 atm). To drop slowly:
- turn off atmospheric pressure device 12 hours before a fill (ie night before). the helium gas should bleed out slowly down (~4 mbar/hr) to ~1atm (980 mbar) now.
- remove helium port stopper slowly and completely and put back top part as a plug.
- remove check valve completely and stopper with kim wipes
- cool down transfer line until you see the flame as usual and insert line
- vent dewar and try and insert line with out o-ring at adapter point. i don't think they are necessary..
- ramp pressure up to 2psi and let fill. takes appx 1.25 hrs from 58%
- The next day come in and clear the inside of the N2 ports from ice build-up. This is a problem on this magnet.
- Also, be sure the green N2 tubing is connected well and the ports are greased.